Inhouse product
Medical face masks are tested against synthetic blood because real blood is hazardous to handle and variable in composition. The Mask Blood Penetration Tester sprays a controlled volume of synthetic blood at the mask surface under pressures that correspond to human blood pressure, then the operator checks whether any liquid has passed through to the inner side. This Mask Blood Penetration Tester is built for mask manufacturers, PPE testing laboratories, and quality inspection departments that need ASTM F2100, ISO 22609, and EN 14683 compliance data without building a custom spray rig.
The Mask Blood Penetration Tester is supplied as a single model, GT-RA01, with the spray assembly, convex sample holder, target plate, and pressure control system configured as a complete benchtop test station.
Model/Part Number: GT-RA01
Applicable Standards: GB 19083 section 5.5, YY/T 0691, YY 0469 section 5.5, ISO 22609, BS EN 14683, ASTM F2100, ASTM F1862
Spraying Distance: 300 to 305 mm (adjustable)
Nozzle Diameter: 0.84 mm
Spraying Speed: 450 cm/s, 550 cm/s, 635 cm/s
Corresponding Pressures: 10.6 kPa, 16 kPa, 21.3 kPa (simulating human blood pressure)
Sample Fixing Device: Convex shape simulating actual mask use, leaving target area intact
Target Plate: Blocks high-pressure edge portion of spray stream, allowing only steady-state flow onto sample
Spray Device: Constant pressure injection, controlled volume over controlled time
Weight: 35 kg
Power Supply: AC 220V, 50 Hz
Mask Blood Penetration Tester main unit
Convex sample fixing device
Constant pressure spray device
Target plate for steady-state flow control
Nozzle assembly with 0.84 mm orifice
Pressure control system for 10.6 kPa, 16 kPa, 21.3 kPa settings
Operation manual
Synthetic blood test fluid (user-supplied to standard specification)
Calibration certificate (available upon request)
Extended warranty (available for purchase)
On-site installation and commissioning (available as a service)
Stable, level benchtop with sufficient space for the 35 kg unit
Dedicated AC 220V, 50 Hz power supply
Compressed air source for the constant pressure spray device (if required by the unit's pneumatics)
Synthetic blood prepared according to the applicable standard's formulation
Adequate ventilation and surface protection for liquid spray testing
Operator training on synthetic blood handling and spray testing procedures is recommended
The Mask Blood Penetration Tester delivers spray speeds of 450 cm/s, 550 cm/s, and 635 cm/s, corresponding to the 10.6 kPa, 16 kPa, and 21.3 kPa pressures specified in ASTM F2100 and ISO 22609. You select the pressure level your standard requires, and the instrument reproduces it without manual adjustment of a hand-built nozzle.
A flat sample holder does not represent how a mask sits on a face. The convex fixing device holds the mask in a curved, in-use geometry and leaves the target area intact for spray exposure. The synthetic blood distributes across the target zone rather than running off a flat surface.
Spray streams have a high-pressure edge that can skew penetration results. The target plate blocks that edge portion and allows only the steady-state flow to reach the sample. This increases the accuracy and repeatability of the liquid velocity that actually contacts the mask surface.
At 35 kg and powered by a standard AC 220V supply, the unit fits on a laboratory bench without a dedicated test room. Mask manufacturers can place it in incoming material inspection or finished goods testing without restructuring the lab layout.
To explore how this Mask Blood Penetration Tester fits into your PPE testing workflow, visit our website at Optimum Solution to compare related mask and blood penetration testing instruments. For detailed application notes on synthetic blood penetration testing, you can view the GESTER mask blood penetration tester product page on the manufacturer's site. For information on medical face mask standards, refer to the ASTM F2100 specification or the ISO 22609 test method.
Struggling to reproduce the same spray pressure across test runs?
The constant pressure spray device holds the selected velocity at 450, 550, or 635 cm/s. You do not need to adjust a manual regulator and guess whether the pressure matches the standard.
Concerned that a flat sample holder is giving unrealistic results?
The convex fixing device holds the mask in a curved geometry that approximates actual use. The target area stays intact, and the synthetic blood distributes across the zone the way it would on a worn mask.
Tired of variable results from spray edge effects?
The target plate blocks the high-pressure periphery and lets only the steady-state core reach the sample. The velocity that contacts the mask is the velocity you set, not a blend of core and edge flow.
Need test data that maps to ASTM F2100 or ISO 22609 review?
The spray speeds and pressure equivalents follow the standard values. The sample geometry and target plate design support repeatable penetration checks, so your records reference a controlled method rather than a bench improvisation.
We provide a 1-year manufacturer's warranty covering parts and labor. Technical support, installation assistance, and operational training are also available. Replacement parts and consumables are available throughout the product's service life.
Request a Quotation or Technical Datasheet: Contact us for more information, customized configurations, and solutions tailored to your specific requirements.
Optimum Solution is a leading supplier of laboratory equipment, analytical instruments, and process control solutions in Bangladesh. Since our inception, we have built a strong presence in the industry, serving research labs, universities, pharmaceutical companies, and quality control laboratories nationwide. Through continuous dedication to improving our products and services, Optimum Solution has earned the trust and loyalty of our valued clients.