Are you looking for a precision autoclave that delivers consistent sterility, high throughput, and absolute safety for your busy laboratory or clinic? Do you need a reliable solution for sterilizing everything from surgical instruments and glassware to culture media, without worrying about failed cycles or equipment damage? The vertical Class N steam autoclave is engineered to meet these exacting demands. Combining a robust stainless steel chamber, microcomputer-controlled automation, and flexible capacity options from 35 liters to 150 liters, this steam sterilizer ensures thorough, reproducible results every time. It is the ideal workhorse for research labs, medical facilities, and pharmaceutical environments where contamination control is critical and workflow efficiency is paramount.
Achieve uncompromising sterility and operational confidence with intelligent features that directly address common lab pain points.
Eliminate Inconsistent Sterilization Results: Fluctuating temperatures or inadequate steam penetration can lead to failed cycles and compromised safety. This laboratory autoclave features a microcomputer-controlled system with a precise sterilization temperature range of 116°C to 134°C. A three-stage temperature rise and fall program enhances steam saturation, ensuring deep penetration even for complex or dense loads, guaranteeing reproducible sterilization outcomes.
Maximize Throughput and Minimize Downtime: Slow sterilization cycles create bottlenecks. With generous chamber volumes ranging from 35L up to 150L, this high-capacity sterilizer accommodates large batches of instruments or bulky media, dramatically improving lab efficiency. The automatic exhaust steam discharge and cooling air circulation help ensure a safe, dry end result with minimal waiting time between cycles.
Ensure Operator and Equipment Safety at All Times: Outdated units can pose serious safety risks. The vertical autoclave is equipped with comprehensive, built-in safety protections, including:
Over-temperature protection
Over-pressure auto discharge
Low water-level protection and anti-dry burn safeguard
Self-expanding seal and safety interlock locking lid
These features prevent accidents, protect valuable equipment, and ensure operator confidence.
Simplify Operation for All Lab Staff: Complex manual controls invite human error. The intuitive digital display and touch-button interface make operation straightforward, reducing reliance on extensive operator experience and minimizing the risk of mistakes during cycle setup.
Engineered for performance, durability, and ease of use:
Robust Stainless Steel Construction: High-quality stainless steel chamber ensures long-term durability and corrosion resistance. Vertical, space-efficient design fits easily into lab or clinic environments.
Automated Microcomputer Control: Precisely set sterilization parameters (temperature and time from 4 to 120 minutes) with minimal manual intervention. The system manages the cycle automatically.
Flexible Capacity Range: Available in multiple volumes (35, 50, 75, 100, 120, 150L) to match your specific throughput needs.
Comprehensive Safety Interlocks: All critical safety features are integrated to protect users and samples.
User-Friendly Interface: Digital display, touch-button controls, and buzzer alert on cycle completion enhance usability.
Product Identity & Core Function
Product Name: Precision Autoclave / Vertical Steam Sterilizer
Type: Class N (Vertical, gravity displacement) Steam Autoclave
Primary Application: Sterilization of instruments, glassware, culture media, liquids, and lab waste in research, clinical, and pharmaceutical laboratories.
Target Users: Microbiology labs, molecular biology labs, hospitals, clinics, biotech facilities, quality control labs.
Chamber & Capacity
Available Chamber Volumes: 35 L, 50 L, 75 L, 100 L, 120 L, 150 L
Chamber Material: High-quality Stainless Steel
Chamber Dimensions (approximate, varies by model):
Smallest (35L): Ø300 × 500 mm
Largest (150L): Ø500 × 780 mm
Included Accessories: Stainless steel basket(s) or bucket(s) for holding loads.
Performance & Operating Parameters
Sterilization Temperature Range: 116°C to 134°C (adjustable)
Sterilization Time Setting Range: 4 to 120 minutes (adjustable)
Design Pressure: 0.25 MPa
Rated Working Pressure: 0.22 MPa
Cycle Program: Three-stage temperature rise/fall for enhanced steam penetration.
Electrical & Power Specifications
Power Supply: AC 220V or 110V, 50/60 Hz (specify voltage when ordering)
Power Consumption: 2.5 kW to 6.0 kW (depending on chamber volume/model)
Dimensions & Weight (Approximate, varies by model)
Example 35L Model:
Outer Dimensions (approx.): 400 × 400 × 860 mm
Gross Weight: ~76 kg
Example 150L Model:
Outer Dimensions (approx.): 605 × 649 × 1210 mm
Gross Weight: ~130 kg
Safety & Control Features
Control System: Microcomputer-controlled cycles with digital display.
User Interface: Touch-button panel with intuitive controls.
Safety Protections:
Over-temperature protection
Over-pressure auto discharge valve
Low water-level protection / Anti-dry burn
Self-expanding door seal
Safety interlock locking lid (prevents opening under pressure)
Cycle Management: Automatic exhaust steam discharge and cooling air circulation. Buzzer alerts on sterilization completion.
Ensure consistent sterility, protect your staff, and maximize throughput. This vertical steam sterilizer delivers the performance and safety your critical work demands.
Optimum Solution is a leading supplier of laboratory equipment, analytical instruments, and process control solutions in Bangladesh. Since our inception, we have built a strong presence in the industry, serving research labs, universities, pharmaceutical companies, and quality control laboratories nationwide. Through continuous dedication to improving our products and services, Optimum Solution has earned the trust and loyalty of our valued clients.