Inhouse product
The AC02 is a water-cooled xenon arc light fastness tester from AVENO, built for simulated light colour fastness, photo-sweat composite colour fastness, and solar radiation aging tests on textiles, rubber, plastics, paints, coatings, automotive materials, and non-metallic products. It uses a 6 kW water-cooled long-arc xenon lamp with a high-transmittance filter system, and irradiance is digitally set, monitored in real time, and automatically adjusted through closed-loop compensation. The chamber controls temperature, humidity, and blackboard temperature in a closed cycle, and the 10.4-inch touchscreen supports multistage programming with online editing. The specifications below cover the standard unit; the AC02-1 and AC02-2 configurations differ in sample capacity.
Model/Part Number: AC02 (AC02-1 and AC02-2 sample capacity variants)
Light Source and Irradiance
Lamp: 6 kW water-cooled long-arc xenon lamp, point light source
Irradiance control range: 0.6–3.08 W/m² at 420 nm
Optional wavelengths: 340 nm, 300–400 nm, or 280–800 nm
Irradiance setting: digital, real-time monitoring, automatic adjustment
Compensation: closed-loop stability automatic compensation
Irradiance sensing: optical fibre guided, 16-bit sampling AD, 32-bit CPU processing
Correction: standard lamp correction technology to eliminate sensor aging attenuation
Lamp regulation: AC variable-frequency stepless regulation
Chamber Conditions
Chamber temperature range: 30–90 °C, ±0.1 °C
Chamber humidity (light cycle): 20–70 %RH, ±0.1 %RH
Chamber humidity (dark cycle): 30–95 %RH, ±0.1 %RH
Standard blackboard thermometer: 35–120 °C, ±0.1 °C
Test time range: ≤ 10,000 h
BPT and BST: same-station equidistant detection with the sample
Sample Handling
Sample frame rotation speed: 3–7 rpm, settable
Sample clamp distribution diameter: Φ550 mm
AC02-1 sample holders: ISO 145 x 45 mm, 33 pcs; AATCC 145 x 70 mm, 21 pcs (option)
AC02-2 sample holders: ISO 145 x 45 mm, 48 pcs; AATCC 145 x 70 mm, 36 pcs (option)
Individual timing: all sample clips timed separately, so samples of different test durations run in the same machine
Control and Cooling
Display: 10.4-inch touchscreen, Chinese and English interface
Monitoring modes: animation, numbers, charts
Programming: multistage running mode, online programming without downtime
Humidification: ultrasonic and PTC
Circulation: servo motor controlled wind speed
Cooling water: internal circulation with high-efficiency heat exchange system
Physical and Electrical
Dimensions: 120 x 90 x 180 cm
Weight: 350 kg
Brand: AVENO
Origin: China
Standards Referenced
Textile: AATCC 169, GB/T 8430, GB/T 14576, GB/T 8427, ISO 105-B02/B04/B06, M&S C09/C09A, JIS L0843
Rubber and plastics: GB/T 16422.1, GB/T 12831, GB/T 9344, ASTM D2565, ISO 4892
Paint coating: GB/T 1865, ASTM D6695, ISO 11341, NES M0135
Non-metallic material: AATCC TM16, ASTM G26, ASTM D4459, ASTM G155, GB/T 2423.24, GB/T 15102, GB/T 15104
Automotive: SAE J2527, SAE J1885, SAE J2212, SAE J1960, SAE J2412, ISO 3917, PV 1303, PV 3929, PV 3930, DVM 0067-MA, HES D6601, TSL 0601G, GMW 3414, EDS-T-7415, D47 1431
Certification
CE
ISO 9001
ISO 17025
Standard Accessories Included:
AC02 base unit with water-cooled lamp holder installed
Xenon arc lamp, 1 pc installed plus 1 pc spare
Sample holders per configuration: 145 x 45 mm ISO/GB holders and 145 x 70 mm AATCC holders
Cover plates for each holder size
Sample boards per configuration
Standard blackboard thermometer, 1 pc
Oscillator for humidifier, 2 pcs
PP cotton, 1 pc installed plus 1 pc spare
Blue wool standard sample, 1 set
Grey scale, 1 set
Inner hexagon spanner, 1 pc
Fuse (5A), 2 pcs
Drainage intake pipe, 1 set
Operating manual
Items Excluded:
Optional irradiance wavelengths (340 nm, 300–400 nm, 280–800 nm) where not ordered
Additional sample holders for non-standard sizes
Calibration certificates for the irradiance sensor
External chiller or water treatment system beyond the internal circulation and heat exchange
Pre-installation Requirement:
Place the tester on a rigid, level floor capable of supporting 350 kg plus water, with clearance on all sides for the chamber door, sample frame access, and the cooling water connections. Connect to a grounded power supply matching the nameplate. Connect the drainage and intake pipes and fill the cooling water circuit before the first run. Confirm the internal circulation and heat exchange system is working before operating the xenon lamp. Allow the chamber to reach stable temperature and humidity before loading samples.
Light fastness testing fails quietly when the light source ages. The xenon lamp's output drops over time, the sensor measuring that output drifts, and the test that once produced a Grade 4 result now produces a Grade 3.5 for reasons that have nothing to do with the sample. The AC02 water-cooled light fastness tester addresses that directly with standard lamp correction technology that eliminates the attenuation of the irradiance acquisition system from long-term sensor aging, and closed-loop automatic compensation that holds the set irradiance through the run.
The chamber control architecture is what makes long tests meaningful. Temperature, humidity, and blackboard temperature are all controlled in a closed cycle, with ultrasonic and PTC humidification and servo-controlled circulation wind speed. For tests that run up to 10,000 hours, the difference between a chamber that holds conditions and one that drifts is the difference between a valid aging result and a curve that cannot be interpreted.
The BPT and BST same-station equidistant detection is a detail that matters for standard conformance. The black panel and black standard thermometers are measured at the same station and distance as the samples, so the recorded blackboard temperature reflects the samples' actual exposure rather than an averaged chamber value. For standards that specify BPT or BST limits, that placement determines whether the test is valid.
For buyers sourcing through OS4U, the configuration questions are the AC02-1 or AC02-2 sample capacity, whether the optional 340 nm or 300–400 nm irradiance bands belong in your methods, and whether calibration certificates should be included. The AVENO AC02 product page lists the full accessory range.
The AC02 water-cooled light fastness tester holds irradiance through closed-loop stability automatic compensation, with digital setting and real-time monitoring. The standard lamp correction technology eliminates sensor aging attenuation, so the irradiance recorded at hour 5,000 means the same thing as at hour 1.
Temperature at 30–90 °C ±0.1 °C, humidity at 20–70 %RH in the light cycle and 30–95 %RH in the dark cycle, and blackboard temperature at 35–120 °C ±0.1 °C. All three are controlled together, not independently, through ultrasonic and PTC humidification and servo-controlled circulation.
Black panel and black standard thermometers are measured at the same station and distance as the sample. The recorded blackboard temperature reflects the sample's actual exposure, which is what the standards that specify BPT and BST limits require.
All sample clips are timed separately. Samples that need different test durations can run in the same machine without waiting for the longest test to finish, which is a practical benefit for labs running multi-standard workloads.
The cooling water uses a mix of internal circulation and high-efficiency heat exchange, which saves water and reduces operating costs across long tests. On a machine that runs for thousands of hours, that matters.
Textile, rubber, plastics, paint coating, non-metallic material, and automotive standards are all referenced, from AATCC 169 and ISO 105-B02 through SAE J2527, SAE J1960, PV 1303, and GMW 3414. One chamber covers the light fastness methods a multi-industry lab is likely to need.
The 10.4-inch touchscreen with animation, number, and chart monitoring modes supports multistage running mode programming with online editing without downtime. Long test sequences can be adjusted without stopping the run.
Getting results that drift between runs? Closed-loop irradiance compensation and standard lamp correction technology hold the set irradiance through the run and eliminate sensor aging drift. The irradiance at hour 5,000 matches hour 1.
Worried about temperature and humidity interacting? Temperature, humidity, and blackboard temperature are controlled in a closed cycle with ultrasonic and PTC humidification. The three conditions hold together rather than drifting independently.
Needing BPT and BST readings that reflect the sample? The black panel and black standard thermometers are measured at the same station and distance as the sample. The recorded value reflects the sample's actual exposure.
Running samples with different test durations? All sample clips are timed separately. Mixed-duration tests run in one machine without waiting for the longest test to finish.
Concerned about water consumption across long tests? The internal circulation with high-efficiency heat exchange saves water across the thousands of hours a light fastness test runs.
Needing to adjust a programme mid-run? Multistage programming with online editing without downtime. The test sequence can be adjusted without stopping the run.
Testing to multiple industry standards? Textile, rubber, plastics, paint, non-metallic, and automotive standards are all referenced. One chamber covers the light fastness methods across industries.
We provide a 1-year manufacturer’s warranty covering parts and labor. Technical support, installation assistance, and operational training are also available. Replacement parts and consumables are available throughout the product’s service life.
Contact us for more information, customized configurations, and solutions tailored to your specific requirements.
Optimum Solution is a leading supplier of laboratory equipment, analytical instruments, and process control solutions in Bangladesh. Since our inception, we have built a strong presence in the industry, serving research labs, universities, pharmaceutical companies, and quality control laboratories nationwide. Through continuous dedication to improving our products and services, Optimum Solution has earned the trust and loyalty of our valued clients.